5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-308

Parameter

Range; Default value / Meaning

Delta

-1e30...1e30; 0

Step size for the variation

This parameter defines the amount by which the selected parameter
with the name

Param.name

is varied. With a positive value of

Delta

, the

parameter is increased, with a negative value it is decreased.

The step size must be entered for all parameters

in the base units s, V, Hz or °.

The following conditions apply to entry of the step size:

Parameter

Unit

Delta

min

Meas.cycles

Number

1

Modul.freq.

Hz

1

Ph.angle

°

1

Prep.cycles

Number

1

Rot.speed

Number/min

200

t.line

s

0.1

t.meas

s

0.0001

t.pulse

s

0.0001

t.step

s

0.00005

U.ampl

V

0.001

U.end

V

0.001

U.meas

V

0.001

U.start

V

0.001

U.step

V

0.002

Description

The

VARNI

instruction defines the variation of a parameter within a VAR

loop. The parameter of the following instruction specified in

Param.name

is increased or decreased after the first execution of the VAR loop by the
amount

Delta

.

Comments

The instruction with the parameter to be varied must be entered after
the

VARNI

instruction.

The parameter to be varied is not increased or decreased by the
amount

Delta

until the second execution of the VAR loop, in the first

execution the value entered in the parameter of the instruction is
used.

Several VARNI instructions for different instructions are possible
within a VAR loop. However, it is not advisable to take advantage of
this possibility to change different parameters simultaneously.

REP loops are allowed within VAR loops, a VAR loop can also be
inserted in a REP loop.

No ADD or CAL loops are allowed within a VAR loop.

Within VAR loops, the

EXIT

instruction can be inserted for a

premature exit from this loop.