303 tsweep, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 454

background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-303

TSWEEP

Page:

S

Stand:

M, A

Abbreviation:

TSW

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. TSWEEP ..... U.meas ........mV

Function

Application of a constant voltage to the electrodes and measurement of
the current with measured value storage.

Parameter

Range; Default value / Meaning

t/s

0.1...999.9 s; -

Measurement duration

The duration of measurement is entered in the line time.

U.meas

-3000...3000 mV; -

Measurement voltage

The voltage entered here is applied to the electrodes. This voltage is
retained even after the line time has elapsed until a SWEEP instruction
follows or until it is switched off by

ØMEAS

or

END

in the main program

on the "OP.SEQUENCE" page.

Description

The

TSWEEP

instruction is used to apply the voltage defined in the

U.meas

parameter to the electrode and start the current measurement in

accordance with the parameters defined in the measurement mode. The
measurement frequency is determined by the time interval

t.step

of the

measurement mode, the measurement duration by the line time of the

TSWEEP

instruction.

In contrast to the

MEAS

instruction, the measured values in measurement

of the current are plotted. It is also possible to display the live curve (see

section 5.5.1), but no evaluation is performed.

I

t

t/s

U

t

U.meas