Metrohm 746 VA Trace Analyzer User Manual
Page 171

5.3 Set Up
746 VA Trace Analyzer / 747 VA Stand
5-20
Parameter
Range; Default value / Meaning
Scan
display only
Input control lines 1...8
Status display of the input control lines 1...8:
0
off, inactive
(5 V)
1
on, active
(0 V)
You will find further details on the control lines in section 3.13.
Set
0,1; 0
Output control lines 1...8
Here, the output control lines 1...8 can be set manually:
0
off, inactive
(open)
1
on, active
(0 V)
You will find further details on the control lines in section 3.13.
Ext. start
on, off; off
External start via remote interface
Enable/disable external start via input control lines 1 and 2 of the
remote interface:
on
External start/stop enabled:
Input control line 1 = on
→
Input control line 2 = on
→
off
External start/stop disabled
You will find further details on the control lines in section 3.13.
Printer
display only
Printer at parallel interface
Title for the printer connected to the 746 VA Trace Analyzer.
Type
Metrohm, IBM Propr., Epson 80*60, Epson 60*72, HP Deskjet,
HP Thinkjet, HP Thinkj.E; none
Printer type
Selection of the printer driver for the external printer attached to the
parallel interface (only for 746.0010 version without built-in printer).
Metrohm
Built-in thermal printer
(only with 746.0020 version)
IBM Propr.
IBM Proprinter and printers with IBM emulation
Epson 80*60
EPSON printers and printers with EPSON emu-
lation with 80
×
60 dpi graphics resolution (24-
pin dot matrix, ink jet and laser printers)
Epson 60*72
EPSON printers and printers with EPSON emu-
lation with 60
×
72 dpi graphics resolution (9-pin
dot matrix, ink jet and laser printers)
HP Deskjet
HP printers (HP DeskJet, HP LaserJet) and
printers with HP PCL3 emulation
HP Thinkjet
HP ThinkJet, HP QuietJet and compatibles
HP Thinkj.E
HP ThinkJet in EPSON mode and compatibles
You will find further details on the attachment of printers and the
required settings in section 3.11.