5.3.1 VA TRACE ANALYZER
746 VA Trace Analyzer / 747 VA Stand
5-17
Softkeys
Meaning
Contrast
down
Decrease screen contrast
The screen is set one level darker (–2). The current contrast value is
displayed in the
Contrast
parameter.
Continuous darkening can be achieved by pressing the keys
Reset
Ifc.1
Reset RS interface 1
RS232 interface 1 is reinitialized. This may be necessary with certain
disturbances in data communication traffic via the RS interface (see
Error messages, section 7.3.2).
Reset
Ifc.2
Reset RS interface 2
RS232 interface 2 is reinitialized. This may be necessary with certain
disturbances in data communication traffic via the RS interface (see
Error messages, section 7.3.2).
Save
Ifc.par.
Storage of RS232 parameters
For both RS interfaces, the five RS232 parameters
Baud rate
,
Data
bits
,
Stop bits
,
Parity
and
Handshake
are not stored definitively
until this softkey has been pressed.
Diagn.
Diagnostic routine
Enter the instrument diagnostic routine (see section 7.6).
Parameter
Range; Default value / Meaning
Program
display only
Program version number
Date
YYYY-MM-DD;
Date
Date in the format year – month – day.
Time
hh:mm:ss;
Time
Time in the format hours – minutes – seconds.
Beeper
on, off; on
Beeper
General switching on/off of the beeper (with
Beeper = off
an audio
signal is put out only with errors occurring during program execution
or with the
BEEP
command).
Display
display only
Screen
Title for those parameters which affect the screen settings.
Switch-off time
15...999 min; 30 min
Switch-off time
Time interval without keypad entry after which the screen will be
automatically blanked.