5.5 Operation

746 VA Trace Analyzer / 747 VA Stand

5-78

Parameter

Range; Default value / Meaning

The display of

no

means that the determination has not been

changed.

Once a determination has been modified, it can no longer be forced to

the unmodified condition

no

. If you wish to keep the determination in

the unmodified condition, it must be stored in the "DETERMINATIONS"

memory or on the "DATA CARD" before the change.

Sample table

display only

Name of the sample data table

Name of the sample data table from which the sample data of the
determination loaded in the working memory have been automatically
fetched (

Auto.samples = on

).

The display of

means that the sample data have not been fetched

from a sample data table (

Auto.samples = off

).

1

st

sample data

line

Input possibilities, see under the individual parameters

Sample data of current sample

Display of the current sample data. The layout of this line is identical
to that of the lines on the "SAMPLES" page (see section 5.4.4). The
meaning of the displayed data and the input possibilities depend on
the instrument status:

In the standby mode

* READY *

, data specific to the sample can

be entered here and can be used for the next determination in
certain cases (see Table).

During an ongoing determination (

* BUSY *

status), the data of

the current sample are displayed and can not be modified at this
time.

What happens to the sample data entered in this line when
is pressed depends on possible entries in the 2

nd

sample data line as

well as on the setting of the two parameters

Auto.samples

and

Auto.batch

:

1

st

line

2

nd

line

Auto.samples

Auto.batch

Determination

filled

blank

off

off

Single determination with
sample data of the 1

st

line

filled or

blank

filled

off

off

Single determination with
sample data of the 2

nd

line

filled

blank

off

on

Series determination with
sample data of the 1

st

line

filled or

blank

filled

off

on

Series determination with
sample data of the 2

nd

line

filled or

blank

filled or

blank

on

off

Single determination with
automatic sample data
transfer from "SAMPLES"

filled or

blank

filled or

blank

on

on

Series determination with
automatic sample data
transfer from "SAMPLES"