5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-198

Example

A measurement with background compensation and standard addition
is normally programmed on the OP.SEQUENCE page with the following
basic structure:

:

DOS>M or DOS/M

Addition of supporting electrolyte

:

to measuring vessel

:

(BACKGND

Start of block for background compensation

SEGMENT Seg1

Call-up of the segment (BACKGND curve)

BACKGND)

End of block for background compensation

:

SMPL>M or SMPL/M

Sample addition to the measuring vessel

(ADD

Loop start for standard additions

(REP

Loop start for replications

SEGMENT Seg1

Call-up of the segment (measured curve)

REP)n

Loop end, n = number of replications

ADD>M, ADD/M or PADD>M

Constant or proportional standard addition

ADD)n

Loop end, n = number of standard additions

:

see also

BACKGND)