6 Data processing and evaluation
746 VA Trace Analyzer / 747 VA Stand
6-32
Ref.subst
Name of reference substance
V0.CSoln.
Addition volume of the calibration solutions
CSoln.name
Name of the calibration solutions
Mass conc.
Mass concentration
ρ
aa
of the calibration solutions
Soln.name
Name of the solution with the internal standard
(for
Mode = referen.
, must be identical with the name
specified in the RADD instruction)
Mass conc.
Mass concentration
ρ
rr
of the solution with the internal stan-
dard (for
Mode = referen.
)
In the content determination using an internal standard, the following procedure is
used to calculate the sought mass concentration
ρ
eff,s
eff,s
of the sample (see also ex-
ample in section 4.4):
1
Measurement of the calibration solutions
The calibration solutions of known mass concentration
ρ
a,n
a,n
and
ρ
r,n
r,n
are each
measured several times. This gives:
EV
n
Evaluation quantity for n
th
calibration solution
(for analyte and reference substance)
EV
n
.mean
Mean value of the evaluation quantities for n
th
calibra-
tion solution (for analyte and reference substance)
Std.dev.
Standard deviation of the n
th
individual value
EV
n
(for measurement and reference substance)
EV
a,n
/
EV
r,n
Ratio of the evaluation quantities for n
th
calibration
solution
ρ
eff,a,n
eff,a,n
Effective mass concentrations of the analyte for the
calibration solutions (calculated from
ρ
a,n
a,n
taking the
dilution into account)
ρ
eff,r,n
eff,r,n
Effective mass concentration of the reference sub-
stance for the calibration solutions (calculated from
ρ
r,n
r,n
taking the dilution into account)
ρ
eff,a,n
eff,a,n
/
ρ
eff,r,n
eff,r,n
Ratio of the effective mass concentrations of analyte
and reference substance for n
th
calibration solution
2
Determination of the calibration curve
The calibration curve is determined using a linear function of type
0/linear
:
y = bx
Straight line through zero point
The parameter b is determined by the procedure described in section 6.9.3
with y =
EV
a
/
EV
r
and x =
ρ
eff,a
eff,a
/
ρ
eff,r
eff,r
by weighted least squares minimization.
This parameter is displayed on the "RESULTS/Calibrations" page and has
the following meaning:
b
Prop.factor
Proportionality factor (slope of the calibration curve)