219 (csweep, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 370

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-219
(CSWEEP
Page:
S
Stand:
M, A
Abbreviation:
(CS
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. (CSWEEP
Function
Start of the block for a cyclic voltage sweep.
Description
The
(CSWEEP
instruction opens a block in a segment in which the ramps
for the cyclic voltage sweep are defined. This block can contain
maximum 5
RAMP
instructions, other instructions are not allowed. The
block must be closed with the
CSWEEP)
instruction.
Comments
•
Only one CSWEEP block is allowed per segment.
•
CSWEEP blocks may not be located within REP loops.
•
The electrode type (admissible are
HMDE
or
RDE
) and the measure-
ment mode
CYCMODE
must be entered before the CSWEEP block.
•
The number of measurement cycles for the CSWEEP block (max. 9)
is defined in the
CYCMODE
instruction by the
Meas.cycles
parameter.
•
Within a CSWEEP block a wait time between two ramps can be
programmed only by insertion of an additional, slow ramp as flat as
possible.
Example
see
RAMP
see also
CSWEEP), CYCMODE, RAMP