292 sqwmode, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 443

background image

5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-292

SQWMODE

Page:

O, S

Stand:

M, A

Abbreviation:

SQ

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SQWMODE ..... U.ampl ........mV Modul.freq. ........Hz

t.step ........s Prep.cycles ........

t.meas ........ms Meas.cycles ........

Function

Measurement mode for square wave voltammetry.

Application

Square wave voltammetry with the

SQWMODE

measurement mode is

primarily suitable for reversible electrode processes. It is used particu-
larly for sensitive stripping voltammetric determinations at the HMDE or
RDE.

Parameter

Range; Default value / Meaning

U.ampl

1...50 mV; 20 mV

Voltage amplitude

Voltage amplitude of the superimposed square wave voltage.

t.step

0.05...9.9 s; 0.3 s

Time interval for voltage step

Time interval after which the voltage in the sweep is increased or
decreased by the amount

U.step

.

The following condition applies to the definition of

t.step

:

Meas.cyc. + Prep.cyc.

t.step > ––––––––––––––––––––– + t.drop + 30 ms

Modul.freq.

t.meas

0.1...32.0 ms; 2.0/1.66 ms for 50/60 Hz

Measurement time

Time during which the current is measured. Measurement is performed
at the end of the time interval

t.step

immediately before the pulse

start and at the end of the pulse.

The following condition applies to the definition of

t.meas

:

0.5 32 ms

––––––––––– – 1 ms > t.meas < –––––––––––

Modul.freq. Meas.cycles

Modul.freq.

10...250 Hz; 50/60 Hz for 50/60 Hz

Modulation frequency

Modulation frequency of the superimposed square wave voltage.

Prep.cycles

0...255; 0

Number of preparation cycles

Number of square wave voltage cycles without current measurement
which are performed for every voltage step before the measurement
cycles.