5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-263

Method PSA_wine SEGMENT

Pb

================================================================================

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

1 DCTMODE t.step 0.30 s t.meas 40.0 ms

2 MEAS 60.0 U.meas -800 mV

3 ØSTIR

4 PSAMODE

5 PSWEEP 11.0 U.start -800 mV Resolution 100 /V

t.depos 5.0 s

6 END

The illustrative program for the determination of lead with PSA runs as
follows:

OPERATION SEQUENCE

Main program

Line 1

Addition of sample

Line 2....4

Switching on the RDE, application of a voltage
of –200 mV

Line 5....6

Addition of HCl and Hg(II) solution

Line 7...13

Preparation of the electrode, regeneration with

CONDC

and formation of the Hg film in the

preparation segment

Prp-800

Line 14...22

Double standard addition for lead determina-
tion

Line 15...19

Replication of the lead measurement in the
segment

Pb

SEGMENT Prp-800

Preparation of the electrode

Line 1....6

Formation of the mercury film

SEGMENT Pb

Determination of lead

Line 1....3

Deposition of lead at –800 mV for 60 s with
stirring

Line 4....6

Deposition of lead at –800 mV for 5 s without
stirring, then PSA measurement

see also

PSAMODE, MEAS, RDE, HMDE, CONDC