5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-302

TSTIR

Page:

O, S

Stand:

M, A

Abbreviation:

TS

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. TSTIR ..... Rot.speed ......../min

Function

Switch on stirrer or a preset line time.

Parameter

Range; Default value / Meaning

t/s

0.1...999.9 s; -

Stirring time

The duration of stirring is entered in the line time.

Rot.speed

0, 200, 400, ..., 3000 /min; 2000 /min

Rotational speed

Setting of the number of revolutions of the stirrer per min.

Description

The

TSTIR

instruction is used to switch on the stirrer of the 747 VA Stand

for the period preset in

t/s

.

Comment

If the stirrer should not be switched off automatically after a definite time,
the

STIR

instruction can also be used.

see also

STIR, ØSTIR