7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-52

7.6.4

Checking RS232 interfaces

For this test a 3.496.8480 Test plug is needed, This plug can be purchased by customers quoting the
above number or they can make it themselves by following the diagram below:

PIN

PIN

PIN

PIN

2

––––––

3

20 ––––––

6

2

––––––

8

4

––––––

5

Connections in 3.496.8480 Plug

1 Prepare instrument for diagnostic test (see section 7.6.2).

2

3

4

5 Plug 3.496.8480 Test plug into 'RS232 Interface 1' connection.

6

7 The test runs automatically.

(If a fault appears, an error message is outputted.)

8

9 Plug 3.496.8480 Test plug into 'RS232 Interface 2' connection.

10

11 The test runs automatically.

(If a fault appears, an error message is put out.)

12 Press several times until the ”MONITORING” page appears.

13 Remove test plug.

7.6.5

Checking control lines

For this test a 3.496.8550 Test plug is needed, This plug can be purchased by customers quoting the
above number or they can make it themselves by following the diagram below:

PIN

PIN

PIN

PIN

1

–––––– 24

5

–––––– 21

2

–––––– 12

9

–––––– 18

3

–––––– 23

10 –––––– 17

4

–––––– 22

11 –––––– 16

Connections in 3.496.8550 Plug