5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-134

Parameter

Range; Default value / Meaning

Display / Plot

display only

Display/plot (for all measurement modes except PSAMODE)

Title for those parameters which concern the display and plotting of
substance curves recorded with the measurement modes DCTMODE,
DPMODE, SQWMODE, AC1MODE or AC2MODE.

I.scale

auto, fixed; auto

Scaling of the current axis

The following two possibilities are available for scaling of the current
axis for the display and plotting of substance curves:

auto

Automatic scaling with matching to the measured
experimental values

fixed

Fixed scaling with the preset values

I.begin

and

I.end

If all measurement values are outside the current axis fixed with

I.scale

, the current axis for printing the substance curve is scaled

automatically. In this case, the message

scale auto

is printed in the

header line.

I.begin

-999.99...999.99 nA...A;

Initial value for scaling of current axis

This parameter appears only if

I.scale = fixed

has been selected.

If no value is entered for

I.begin

, the initial value is determined

automatically.

I.end

-999.99...999.99 nA...A;

Final value for scaling of current axis

This parameter appears only if

I.scale = fixed

has been selected.

If no value is entered for

I.end

, the final value is determined auto-

matically.

U.div

10 mV/cm...50 V/cm; 50 mV/cm

Division of the voltage axis

Division of the voltage axis in mV/cm for the plotting of substance
curves (not effective in display of the curves).

U.begin

-3000...3000 mV;

Initial value for scaling of voltage axis

If no value is entered for

U.begin

, the initial voltage

U.start

of the

sweep is taken as the initial value.

U.end

-3000...3000 mV;

Final value for scaling of voltage axis

If no value is entered for

U.end

, the final voltage

U.end

of the sweep

is taken as the final value.

I.end

I.begin

U.begin