5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-241
Description
The
FSWEEP
instruction (Fast Sweep) starts the voltage sweep defined in
the parameters. During the sweep, the potentiostat always operates in
the operating mode "fast".
U
t
U.start
U.end
U.step
t.step
Comments
•
The electrode type must be defined before the
FSWEEP
instruction.
Admissible are
DME
,
SMDE
,
HMDE
or
RDE
.
•
The measurement mode must be defined before the
FSWEEP
instruction. Admissible are
DCTMODE
,
DPMODE
,
AC1MODE
,
AC2MODE
or
SQWMODE
, but not
PSAMODE
or
CYCMODE
.
•
The use of the
FSWEEP
instruction is practical only with the measure-
ment modes
DPMODE
and
SQWMODE
. With the other allowable meas-
urement modes,
FSWEEP
is identical to
SWEEP
.
•
The
FSWEEP
instruction should always be used instead of the
SWEEP
instruction with adsorptive stripping techniques (e.g. Ni and Co with
DMG).
•
With the
FSWEEP
instruction maximum 1000 measured values can be
recorded.
•
Apart from the
FSWEEP
instruction, no other SWEEP instruction is
allowed in the same segment.
•
If the final voltage
U.end
should not be switched off at the end of the
program, enter a suitable standby voltage
U.standby
in a
ØMEAS
instruction.
see also
SWEEP, DSWEEP, SSWEEP, DME, SMDE, HMDE, RDE, DPMODE,
SQWMODE