5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-140
Parameter
Range; Default value / Meaning
Evaluation
display only
Evaluation
Title for those parameters which concern the evaluation of sub-
stances.
Mode
VA, PSA; VA
Evaluation mode
Selection between evaluation mode for measurements with
PSAMODE (
PSA
) and those for all other VA measurement modes (
VA
).
Substances can be evaluated only if all substance pa-
rameters and program instructions correspond to the
selected evaluation mode.
Quantity
I.peak, P.peak; I.peak
Evaluation quantity
Whereas only the wave height
I.wave
can be determined with waves
(DCTMODE with SWEEP), there are two different evaluation quantities
with peaks:
I.peak
Peak height (current I in A)
P.peak
Peak area (power P in W)
In the case of a linear baseline with
Scope = f.half
or
Scope = r.half
, only the area up to the peak maximum
is taken as the evaluation quantity.
I.peak
I.wave
P.peak
Scope = whole
Scope = f.half
Scope = r.half
This parameter is not displayed for substances recorded with meas-
urement mode PSAMODE.
Sign. digits
1...5; 4
Significant digits
Number of significant figures for the calculated mass concentration
Mass. conc.
on the "RESULTS/Evaluations" page.