5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-140

Parameter

Range; Default value / Meaning

Evaluation

display only

Evaluation

Title for those parameters which concern the evaluation of sub-
stances.

Mode

VA, PSA; VA

Evaluation mode

Selection between evaluation mode for measurements with
PSAMODE (

PSA

) and those for all other VA measurement modes (

VA

).

Substances can be evaluated only if all substance pa-

rameters and program instructions correspond to the

selected evaluation mode.

Quantity

I.peak, P.peak; I.peak

Evaluation quantity

Whereas only the wave height

I.wave

can be determined with waves

(DCTMODE with SWEEP), there are two different evaluation quantities
with peaks:

I.peak

Peak height (current I in A)

P.peak

Peak area (power P in W)
In the case of a linear baseline with

Scope = f.half

or

Scope = r.half

, only the area up to the peak maximum

is taken as the evaluation quantity.

I.peak

I.wave

P.peak

Scope = whole

Scope = f.half

Scope = r.half

This parameter is not displayed for substances recorded with meas-
urement mode PSAMODE.

Sign. digits

1...5; 4

Significant digits

Number of significant figures for the calculated mass concentration

Mass. conc.

on the "RESULTS/Evaluations" page.