7.1 Electrical safety

746 VA Trace Analyzer / 747 VA Stand

7-1

7 Safety, Errors,

Troubleshooting,

Diagnosis, GLP

7.1

Electrical safety

While electrical safety in the handling of the 746 VA Trace Analyzer and 747 VA

Stand is assured in the context of the specifications IEC 1010-1 (protection class I,

degree of protection IP20), the following points should be noted:

Mains connection

Setting the mains voltage, checking the mains fuse and the mains

connection must be effected in accordance with the instructions in

section 3.2.

Opening the instruments

When the 746 VA Trace Analyzer and 747 VA Stand are connected to

the power supply, neither the 746 VA Trace Analyzer not the 747 VA

Stand or the 695 Autosampler may be opened nor parts of them be

removed, otherwise there is a danger of coming into contact with

components which are live. Always disconnect the instruments from

all voltage sources before you open it and ensure that the mains

cable is disconnected from mains connection plug 13

13 !

Protection against static charges

Electronic components are sensitive to static charging and can be

destroyed by discharges. Before you touch any of the components

inside the 746 VA Trace Analyzer or 747 VA Stand, you should earth

yourself and any tools you are using by touching an earthed object

(e.g. green housing of the instrument or a radiator) to eliminate any

static charges which exist.