5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-261

PSWEEP

Page:

S

Stand:

M, A

Abbreviation:

PSW

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. PSWEEP XXXXX U.start ........mV Resolution ......../V

t.depos ........s

Function

Sweep for

PSAMODE

measurement mode.

Parameter

Range; Default value / Meaning

t/s

display only

Time for sweep

The time for the entire sweep is displayed in the line time (which can
not be edited) and is calculated as follows:

Sweep time = t.depos + 6 s

U.start

-3000...3000 mV; -

Initial voltage

Initial voltage for the sweep. This voltage is applied to the electrodes
during the enrichment time

t.depos

.

t.depos

0.1...600.0 s; 60.0 s

Deposition time

During this time, the voltage

U.start

is applied to the electrodes. The

preset voltage is then switched off and the free voltage drop measured
for 6 s.

Resolution

50...150 /V; 100 /V

Resolution

Number of intervals per volt for conversion of the segment curves U vs t
into substance curves t/U vs U. The lower the resolution selected, the
greater the smoothing of the substance curve.

Description

The

PSWEEP

instruction starts the sweep defined in the parameters for

the

PSAMODE

measurement mode. A constant voltage

U.start

is

applied to the electrodes for a predetermined time

t.depos

, the freely

dropping voltage U is then measured for 6 s as a function of time.

Adjustable parameters:

PSWEEP

U.start
t.depos
Resolution

U

t

U.start

t.depos

t.meas = 6 s