beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 296

background image

5.6.3 SUBSTANCES

746 VA Trace Analyzer / 747 VA Stand

5-145

Parameter

Range; Default value / Meaning

Coefficients

display only

Coefficients

Title for coefficients of the standard addition calculation.

Y.reg

display only

Regression value calculated for sample

Display of the value

Y.reg

in A, W or s.

Y.reg

is the evaluation quantity (current or area) for the sample

concentration

ρ

eff,s

calculated from the standard addition curve (see

also

Curve type

).

This value can be used in formulae as

KY:

(

= sub-

stance name) for further calculations.

Slope

display only

Slope of standard addition curve

Display of the slope in AL/g, WL/g or sL/g.

With non-linear curves

Slope

represents the slope in the linear region

(see also

Curve type

).

This value can be used in formulae as

KS:

(

= sub-

stance name) for further calculations.

Nonlin.

display only

Non-linearity factor

Display of the non-linearity factor for a non-linear standard addition
curve in [Al/g]

4

, [WL/g]

4

or [sL/g]

4

(see also

Curve type

).

This value can be used in formulae as

KN:

(

= sub-

stance name) for further calculations.

Mean dev.

display only

Mean deviation of standard addition curve

Display of the calculated, mean deviation of the measured values
about the standard addition curve in A, W or s.

The size of

Mean dev.

can be used as a measure of the quality of a

standard addition.

This value can be used in formulae as

KD:

(

= sub-

stance name) for further calculations.

Additions

display only

Additions

Title for the standard addition solutions. In conjunction with CASE
instructions, maximum 4 different standard addition solutions can be
defined.