5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-287

SMDE

Page:

O, S

Stand:

M, A

Abbreviation:

SM

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SMDE ..... Drop size ........

Function

Switch on Static Mercury Drop Electrode (SMDE).

Application

The SMDE is primarily used for sensitive measurements in which the
surface of the mercury drop must be renewed for every measurement.

Parameter

Range; Default value / Meaning

Drop size

1...9; 4

Drop size

Setting of the size of the mercury drop. The area of the drop is
ca. 0.15 mm

2

(

1

) ... 0.60 mm

2

(

9

).

The set drop time

t.drop

important for certain parameter settings is

defined for the SMDE as follows:

t.drop = Drop.size × 40 ms

Description

The

SMDE

instruction switches on the 6.1246.020 Multi-Mode Electrode

as a static mercury drop electrode. The SMDE combines the features of
the DME and HMDE: as with the DME, the drops are constantly
renewed, but during the measurement the drop area is constant as in the
HMDE case. In a subsequent voltage sweep, the Hg drops are knocked
off by the tapping mechanism after the time

t.step

set in the measure-

ment mode.

t.drop

t.step

t

F (Hg)

Area

(Hg)