beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 290

background image

5.6.3 SUBSTANCES

746 VA Trace Analyzer / 747 VA Stand

5-139

Parameter

Range; Default value / Meaning

f.double

r.double

Wave-shaped curves recorded with DCTMODE are evalu-

ated only with

Scope = whole

.

dU.front

auto, 10...500 mV; auto

Difference to front base point

Absolute voltage difference between base point at the peak/wave front
half and the peak voltage

U.peak

or halfwave voltage

U/2

.

With

dU.front = auto

, the difference is determined automatically

(default setting).

S.front

auto, -99.99...99.99; auto

Slope at front base point

Slope of the baseline at its start point.
With

S.front = auto

, the slope is determined automatically (default

setting).
For peak baselines of the linear type, the slope is always calculated
automatically, any numeric entry is ignored.
For baselines of the

AC2 rel.

and

AC2 abs.

types, the slope is not

needed, any numeric entry is ignored.

dU.rear

auto, 10...500 mV; auto

Difference to rear base point

Absolute voltage difference between base point at the peak/wave rear
half and the peak voltage

U.peak

or half-wave voltage

U/2

.

With

dU.rear = auto

, the difference is determined automatically

(default setting).

S.rear

auto, -99.99...99.99; auto

Slope at rear base point

Slope of the baseline at its end point.
With

S.rear = auto

, the slope is determined automatically (default

setting).
For peak baselines of the linear type with

Scope = whole

or

Scope =

f.double/r.double

, the slope is always calculated automatically, any

numeric entry is ignored.
For baselines of the

AC2 rel.

and

AC2 abs.

types, the slope is not

needed, any numeric entry is ignored.

S.rear

S.front

dU.front dU.rear

Base point

(rear)

Base point

(front)