5.6.3 SUBSTANCES
746 VA Trace Analyzer / 747 VA Stand
5-139
Parameter
Range; Default value / Meaning
f.double
r.double
Wave-shaped curves recorded with DCTMODE are evalu-
ated only with
Scope = whole
.
dU.front
auto, 10...500 mV; auto
Difference to front base point
Absolute voltage difference between base point at the peak/wave front
half and the peak voltage
U.peak
or halfwave voltage
U/2
.
With
dU.front = auto
, the difference is determined automatically
(default setting).
S.front
auto, -99.99...99.99; auto
Slope at front base point
Slope of the baseline at its start point.
With
S.front = auto
, the slope is determined automatically (default
setting).
For peak baselines of the linear type, the slope is always calculated
automatically, any numeric entry is ignored.
For baselines of the
AC2 rel.
and
AC2 abs.
types, the slope is not
needed, any numeric entry is ignored.
dU.rear
auto, 10...500 mV; auto
Difference to rear base point
Absolute voltage difference between base point at the peak/wave rear
half and the peak voltage
U.peak
or half-wave voltage
U/2
.
With
dU.rear = auto
, the difference is determined automatically
(default setting).
S.rear
auto, -99.99...99.99; auto
Slope at rear base point
Slope of the baseline at its end point.
With
S.rear = auto
, the slope is determined automatically (default
setting).
For peak baselines of the linear type with
Scope = whole
or
Scope =
f.double/r.double
, the slope is always calculated automatically, any
numeric entry is ignored.
For baselines of the
AC2 rel.
and
AC2 abs.
types, the slope is not
needed, any numeric entry is ignored.
S.rear
S.front
dU.front dU.rear
Base point
(rear)
Base point
(front)