Metrohm 746 VA Trace Analyzer User Manual
Page 204

5.4.2 DETERMINATIONS
746 VA Trace Analyzer / 747 VA Stand
5-53
Softkeys
Meaning
>>
Switch to additional softkeys
all
Print list of all determinations
The " DETERMINATIONS" page is printed out with all determination
entries.
Select
all
Select all determinations
All determinations on the page are selected and marked by an arrow
–
>
.
Recalc.
file(s)
Recalculate selected file(s)
All determinations selected and marked by an arrow
–>
are recalcu-
lated with the method currently loaded in the working storage. For
each of the selected determinations, a recalculation
from original
and a recalculation
from Op. Sequence
(see softkey
on
the "MONITORING" page, section 5.5.1) using the originally recorded
experimental data and the modified method parameters is performed.
The result of the recalculation is seen on the reports and/or curves
defined on the "DOCUMENTATION" page which are automatically put
out to the printer or RS interfaces.
Make sure that only determinations are selected for recalculation which
were recorded with a method whose
name
Method
is identical to the
method name of the method currently loaded in the working storage.
The result of the recalculation is not stored automatically in the deter-
mination file.
Secure/
Unsecure
Secure/unsecure selected determination
The determination selected with the cursor is secured (marking
*
appears on the left next to the determination name) or unsecured
(marking
*
is deleted).
Secured determinations can not be deleted or overwritten.
Change
full cd.
Specify mode when memory full
This softkey can be used to switch between the two possibilities
push
out
or
stop
for the
Full condition
parameter (see under Parame-
ters).
Sort
Sort determinations alphabetically
The determinations are sorted alphabetically. Here, numbers take
precedence over letters, uppercase letters precedence over lower-
case letters.