7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-24

Error message

Meaning/Causes

Rectification

Exit

SAMPLER - ERROR 16:
Siphoning volume XXX
too low

There is too little solution for

the desired transfer com-

mand at the vessel XXX on

the sample rack (SXX), on

the standard racks (AXX,

BXX) or in the vessel with

cleaning solution (C01; soft-

key

) of the 695

Autosampler

Liquid sensor dirty

Check vessel or reduce transfer

volume

V.add

Clean liquid sensor

E/N

SAMPLER - ERROR 17:
Incorrect air pump
signal

Liquid sensor dirty

Air pump for liquid sensor

faulty

Clean liquid sensor

Inform Metrohm service

E/S

SAMPLER - ERROR 18:
Illegal X/Y position

695 Autosampler incorrectly

adjusted

Inform Metrohm service

E/S

SAMPLER - ERROR 19:
Illegal X or Y position

695 Autosampler incorrectly

adjusted

Inform Metrohm service

E/S

SAMPLER - ERROR 20:
Missing 717 Sample
Changer

Connection to 717 Sample

Changer broken:

connecting cable not

plugged in

wrong address

Check connection to 717 Sample

Changer:

plug in connecting cable

restart 746 VA Trace Analyzer

E/S

SAMPLER - ERROR 21:
717 no rack detected

No rack mounted on 717

Sample Changer

Mount rack

E/S

SAMPLER - ERROR 22:
717 overload

Motor at 717 Sample Changer

blocked

Check 717 Sample Changer and restart

E/S

SAMPLER - ERROR 23:
717 no beaker detected

No beaker on the sample rack

of the 717 Sample Changer

Load sample beakers on sample rack

E/S

SAMPLES - ERROR 1:
Missing sample size/S0

No value for

Sample size/S0

available

Enter numeric value or

*

for

Sample

size/S0

on the ”SAMPLES” page

E/C

SAMPLES - ERROR 2:
Blank sample line found

Inadmissible blank line on the

”SAMPLES” page

Delete or fill in blank line

E/C

SAMPLES - ERROR 3:
Method in Meth.call not
found

The method called up with

Method.call

is missing in the

method memory

Load method into method memory or

change method name

Method.call

E/C

SAMPLES - ERROR 4:
Inconsistent S2.Var

No density information (if

sample quantity entered as

weight,

S2

must be entered as

density in g/mL)

Enter density in g/mL under

S2

E/C

SAMPLES - ERROR 5:
Equal entry already
exists

The sample table contains two

lines with identical entries

Check sample data on the ”SAMPLES”

page and change

E/C

SAMPLES - ERROR 6:
Can not reconfig. while
RS232 busy

The data receipt from a

balance can not be switched

on if the RS interface is still

active

Wait a short time then press

(retry) and try again

If the error reappears, press

(abort) and reset RS interface with the

softkey

on the ”VA

TRACE ANALYZER” page

r/a

SEGM. - ERROR 1:
Missing END instruction

Missing

END

instruction

Append

END

instruction to end of

program in the segment

Esc

SEGM. - ERROR 2:
Nesting violation: line
##

Error in the nesting of instruc-

tions with brackets

Check program in the segment

for correct use of instructions with

brackets and correct if necessary

Esc

SEGM. - ERROR 3:
Incorrect bracket
matching: line ##

Wrong use of instructions with

brackets (e.g.

(ADD

.....

REP)

)

Check program in the segment

for correct use of instructions with

brackets and correct if necessary

Esc