7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-11
Error message
Meaning/Causes
Rectification
Exit
DATA CARD - ERROR 9:
File name not found
Error in copying from or to data
card
−
Attempt copying again
−
If error reappears, delete file in
question
−
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 10:
File already closed
Error in copying from or to data
card
−
Attempt copying again
−
If error reappears, delete file in
question
−
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 11:
File already opened
Error in copying from or to data
card
−
Attempt copying again
−
If error reappears, delete file in
question
−
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 12:
Card is already full
No more free storage space on
data card
Use new data card or clear data
E/C
DATA CARD - ERROR 13:
Invalid mode
Error in copying from or to data
card
−
Attempt copying again
−
If error reappears, delete file in
question
−
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 14:
File is protected
The selected file can not be
copied to the data card as a
protected file with the same
name is already present
(overwrite protection)
Disable overwrite protection with
or change file
name
E/C
DATA CARD - ERROR 15:
Unexpected end of file
Error in copying from or to data
card
−
Attempt copying again
−
If error reappears, delete file in
question
−
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 16:
Card is write protected
The data card is write protected
Disable overwrite protection or use a
different data card
E/C
DATA CARD - ERROR 17:
Undefined error
Unspecified error in data traffic
with the data card
Press Reset key
E/C
DATA CARD - ERROR 18:
Bad battery charge
detected
Charge of data card battery
bad
−
Copy all data on data card to instru-
ment memory
−
Change battery
−
Recopy data to data card
E/C
DATA PROC - ERROR 1:
Not enough measuring
points
Too few measured points
available for the smoothing
Extend measurement range or reduce
U.step
E/C
DATA PROC - ERROR 2:
Current overload, data
not evaluable
More than 2 overflow measured
values have appeared in
succession, the curve can thus
not be evaluated
Start new measurement or remove
overflow measured values manually
using the spike elimination
E/C
DATA PROC - ERROR 3:
No background sweep
found
No background sweep avail-
able
Rerecord measurement with back-
ground compensation
E/C
DIAGNOSIS - ERROR 1:
No VA Stand found
Diagnostic fault; connection to
747 VA Stand broken:
−
connecting cable not
plugged in
−
wrong address
Check connection to 747 VA Stand:
−
plug in connecting cable
−
set address correctly and restart 746
VA Trace Analyzer
E/C