5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-162

Parameter

Range; Default value / Meaning

Instrument variables
The following instrument variables can be used in formulae:

C##

Common variable C01...C18 from the working
memory (see section 5.4.5). The common vari-
ables can be used to exchange results between
different determinations.

ED::

Difference of the mean values of the evaluation
quantities (

I.delta

,

P.delta

or

t.delta

) for the

specified substance

and variation

.

The numeric value put out refers to the basic
unit A, W or s.

ES::

Standard deviation of evaluation quantities
(

Std.dev.

) for the specified substance

and variation

. The numeric value put out

refers to the basic unit A, W or s.

EU::

Mean value of peak voltages (

U

) or half-wave

voltages (

U/2

) for the specified substance

and variation

. The numeric value

put out refers to the basic unit V.

EV::

Mean value of evaluation quantities (

I.mean

,

P.mean

or

t.mean

) for the specified substance

and variation

. The numeric value

put out refers to the basic unit A, W or s.

MA:

Standard addition mass (

Add.mass

) for the

specified substance

. The numeric

value put out refers to the basic unit g.

MC:

Mass concentration

ρ

s

(

Mass conc.

) for the

specified substance

. The numeric

value put out refers to the basic unit g/L.

MD:

Total scatter (

MC.dev.

) of mass concentration

ρ

s

for the specified substance

. The nu-

meric value put out refers to the basic unit g/L.

ML:

Scatter due to calibration (

Cal.dev.

) of mass

concentration

ρ

s

for the specified substance

. The numeric value put out refers to the

basic unit g/L.

MS:

Substance mass (

Mass

) in the measuring vessel

for the specified substance

. The nu-

meric value put out refers to the basic unit g.

KD:

Mean scatter (

Mean deviat.

) for calibration and

standard addition curve of the specified sub-
stance

. The numeric value put out re-

fers to the basic unit A, W or s.

KN:

Non-linearity factor (

Nonlin.

) for calibration and

standard addition curve of the specified sub-
stance

. The numeric value put out re-

fers to the basic unit [AL/g]

4

, [WL/g]

4

or

[sL/g]

4

.