5.7.1 General guidelines

746 VA Trace Analyzer / 747 VA Stand

5-175

5.7

Program instructions

5.7.1

General guidelines

The 746 VA Trace Analyzer can be used for the free programming of the sequence

of voltammetric determinations with the aid of program instructions, the associated

parameters and a line time which can be set to any value. The two following dialog

pages are available for the entry of these instructions:

"OP.SEQUENCE"

Main program with call-up of the segments

"SEGMENTS"

Max. 8 segments for measured value recording

(sweeps, one sweep allowable per segment) or macros

Each program instruction is located on an instruction line with the following basic

structure:

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. DCTMODE ..... t.step ........s t.meas ........s

The following rules apply to the basic elements of the instruction line (see also sec-

tion 5.6.1 and section 5.6.2):

Line number

The lines are numbered consecutively on entry of the instruc-

tions. Maximum 99 instruction lines are possible on the

"OP.SEQUENCE" page, maximum 29 instruction lines per

segment on the "SEGMENTS" page.

Line attribute

Each instruction line can be optionally provided with one of the

following attributes:

C

Calibration; the marked instruction line is executed only

when the program run mode is set to

Run mode =

calibration

D

Determination; the marked instruction line is executed

only when the program run mode is set to

Run mode =

determination

.

!

Inactive line; this line is not executed.

@..@

Inactive range; the range between the two lines

marked by

@

is not executed.

Auxiliary parameters

Line
time

Program
instruction

Main parameters

Line attribute

Line number