beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 493

background image

6 Data processing and evaluation

746 VA Trace Analyzer / 747 VA Stand

6-30

m

4

Meas.dev.

is calculated from

σ

x,M

x,M

(see section 6.9.3).

m

2

If sufficient calibration points are available (see above),

Meas.dev.

is estimated from

Cal.dev.

and the message

dev.estimated

put out on the "RESULTS/Evaluations" page.

If this is not the case,

Meas.dev.

is calculated from

σ

x,M

x,M

(see

section 6.9.3).

m = 1

If sufficient calibration points are available (see above),

Meas.dev.

is estimated from

Cal.dev.

and the message

dev.estimated

put out on the "RESULTS/Evaluations" page.

If this is not the case,

Meas.dev.

=

0

is set and the message

no meas.deviation

put out.

The result determination with the aid of a calibration curve saves time compared

with standard additions, but is reliable only

– if the matrix of all samples and calibration solutions is identical or has

no influence on the measurement

– if all measurement parameters (capillary, temperature, etc.) remain un-

changed during measurements

– if the accuracy of the results obtained is checked regularly with the

standard addition method.

With regard to optimum accuracy and scatter, a number of rules must be observed

with calibration curves:

Check linearity range

In development of the method, the linearity range of the calibration curve

should be checked for each substance by recording the curve over a wide

concentration range. Using the calibration curve shown on the "CURVES"

page, you can then determine the region in which the curve is linear and that

in which it is nonlinear.

Working in the linear range

If you wish to perform determinations only in the linear range, to keep the

scatter as low as possible it is advisable to calibrate above all in the lower

and upper part of this range and select as many replications as possible.

Working in the nonlinear range

If you wish to perform determinations in the nonlinear range, as many calibra-

tion solutions as possible with concentrations uniformly distributed over the

entire range should be recorded so that the nonlinear curve shape can be

determined exactly. To keep the scatter as low as possible, select as many

replications as possible.

Checking the offset

Before you deliberately force the calibration curve through the zero point, you

should first check curves with

linear

or

nonlin

to see whether the offset

Y.offset

may be neglected. The size of the offset indicates a possible sys-

tematic error or blank value. To convert this error into the effective mass con-

centration in g/L,

Y.offset

must be divided by

Slope

.

Determining the working range

The calibration curve is defined only for the range between the calibration

solutions with the lowest and highest mass concentrations. Extrapolations

outside this range are not allowed. Hence, enter the maximum possible