3 rs232 interfaces, Features of the rs232 interfaces -10, Data transmission protocols -11 – Metrohm 746 VA Trace Analyzer User Manual
Page 571: Settings and modes -11, Key remote control mode -12, Handshake -14, Pin assignment -17

8 Technical data
746 VA Trace Analyzer / 747 VA Stand
8-10
8.3
RS232 interfaces
This section describes the technical data and features of the RS232
interfaces 16
16 and 17
17 of the 746 VA Trace Analyzer. You will find details
regarding attachment of instruments to these interfaces in section
3.12.
8.3.1
Features of the RS232 interfaces
The two RS232 interfaces of the 746 VA Trace Analyzer are configured as DTE
(Data Terminal Equipment) with the following technical specifications:
Standard
Data interface in accordance with EIA standard RS 232C
(DIN 66020, page 1), transmission parameters can be
set on dialog page "VA TRACE ANALYZER", see section
5.3.1).
Code
8 Bit World Trade Character Set 437 (USA)
Baud rate
300, 600, 1200, 2400, 4800, 9600 (adjustable)
Handshake
HWshort, HWfull, SWchar, SWline, none (details, see
section 8.3.5)
Parity
even, odd, none
Stop bits
1 or 2
Data bits
7 or 8
Max. line length
80 characters + CR LF
Control characters
CR
Carriage Return
DEC 13 HEX 0D
LF
Line Feed
DEC 10 HEX 0A
XON
Data transmission start DEC 17 HEX 11
XOFF Data transmission stop DEC 19 HEX 13
SOH
Start of Heading
DEC 01 HEX 01
ETX
End of text
DEC 03 HEX 03
EOT
End of transmission
DEC 04 HEX 04
CAN
Cancel
DEC 24 HEX 18
Mode
Full duplex (simultaneous transmission and receiving)
Restriction:
If data are first received by the interface, transmission is
not started until the receipt is at an end.
Cable
For interconnection of the 746 VA Trace Analyzer with
third-party devices, only a shielded data cable may be
used (e.g. Metrohm D.104.0201). The cable shield must
be perfectly earthed at both devices (pay attention to
current loops: always use star earthing). Use only
connectors with sufficient shielding (e.g. Metrohm
K.210.0001 with K.210.9004).
Cable length
max. ca. 15 m