Metrohm 746 VA Trace Analyzer User Manual
Page 50

3.4 Multi-mode electrode (MME)
746 VA Trace Analyzer / 747 VA Stand
3-23
3.4.7
Replenishing the mercury (without changing capillary)
The multi-mode electrode 38
38 can also be refilled with mercury without having to
remove the glass capillary 109
109.
1 Dismantle multi-mode electrode
•
Unscrew FEP tubing 51
51 and 59
59 from the MME. Disconnect electrode
cable 37
37 from the MME.
•
Take multi-mode electrode 38
38 out of the measuring head 41
41 and tap the
MME gently to knock off any mercury drops on the glass capillary into the
measuring vessel.
•
Place multi-mode electrode 38
38 horizontally in the electrode holder 114
114
(see Fig. 15-3). The slotted screw 101
101 is now at the top.
2 Replenish mercury
•
Unscrew slotted screw 101
101 using a suitable coin. If the slotted screw 101
101
can not be loosened by hand, screw on locking ring 95
95 and pull out of the
MME.
•
Draw up mercury using the syringe 111
111 with attached needle 112
112 and
expel into the Hg reservoir 103
103.
The Hg reservoir 103
103 must never be filled more than
2
/
3
full with
mercury.
•
Reinsert slotted screw 101
101 into screw thread 100
100 and screw flush to
surface using a suitable coin (this action may expel a few drops of mer-
cury from glass capillary 109
109).
Do not turn so tightly that the cemented-in steel threaded ring 100
100
becomes loose and hence jeopardizes the tightness and safety of the
MME!