5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-258
Page:
O, S
Stand:
M, A
Abbreviation:
PR
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. PRINT ..... ....................................................
Function
Printout of a message.
Parameter
Range; Default value / Meaning
Remark
52 characters; -
Message which is printed out
The comment entered here is printed out on the printer followed by CR
and LF.
All numbers, letters and characters available on the keypad can be
printed out, in addition the entry of all ASCII characters with decimal
code 032
…
255 is possible (see section 5.2.3). Printout of the special
characters with decimal code 000
…
031 is not allowed as they are used
as control characters for the printer.
Description
The
instruction is used to put out the message entered in the
Remark
parameter on the built-in thermal printer or the external printer
connected to the parallel interface.
Comments
If a message should be printed out on an external printer attached to one
of the two RS232 interfaces, the
SEND>RS1
or
SEND>RS2
instructions
must be used.
see also
SEND>RS1, SEND>RS2