5.5.1 MONITORING

746 VA Trace Analyzer / 747 VA Stand

5-79

Parameter

Range; Default value / Meaning

In summary, the data entered in the 1

st

sample data line are used only

for the next determination

if the automatic sample data transfer is switched off
(

Auto.samples = off

), and

if no sample data are entered in the 2

nd

line.

2

nd

sample data

line

Input possibilities, see under the individual parameters

Sample data of sample being prepared

The layout of this line is identical to that of the lines on the "SAMPLES"
page (see section 5.4.4). The meaning of the displayed data and the
input possibilities depend on various factors:

With methods without sample preparation cycles, here both in the
standby mode

* READY *

and during an ongoing determination

(

* BUSY *

status) data specific to samples can be entered and

used for the next determination (see Table) when the automatic
sample data transfer (

Auto.samples = off

) is switched off.

In the case of methods with sample preparation cycles (possible
only with 695 Autosampler and

Auto.batch = on

), here the data

of the sample being prepared are shown and can not be
changed.

Pos.

1...160;

Position number

Position number on the sample rack of the 695 Autosampler (without
695 Autosampler this column has no meaning, however it can be
used for simple sample numbering).

The following rules apply to the position numbering:

All instructions concerning the sample vessel (FEED) (e.g.

SMPL/F

,

CSOLN/F

,

F>M

,

DOS>F

, etc.) are referred to the vessel at

the rack position

Pos.

defined by the position number.

All instructions concerning the auxiliary vessel (AUX) (e.g.

F>A

,

DOS>A

, etc.) are referred to the vessel which immediately follows

the sample vessel defined by the position number (rack position

Pos. + 1

).

The sequence of the positions can be freely selected for sample
solutions, but the same position number can only be entered
once.

With calibration solutions, all solutions must be entered in imme-
diate succession, however, the position numbers themselves can
be selected at random.

During series determinations without automatic transfer of sample
data

Auto.samples = off

, the position number of the current

sample is shown on the 1

st

sample data line in the parameter

Pos.

.

With the UV sample racks, the position numbers on the sample
holders are defined as follows:

6.2041.100 Sample Rack: Sample holder left

1...12

Sample holder right

13...24

6.2041.110 Sample Rack: Sample holder left

1...16

Sample holder right

17...32