5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-188

Description

In the

AC2MODE

measurement mode, a sinusoidal alternating voltage with

a small, constant amplitude and low frequency is superimposed on a
stepwise rising direct voltage ramp. The 2

nd

harmonic wave of the alter-

nating current component I produced by the alternating voltage is
measured as a function of the voltage U. The nature of the measurement
is illustrated below using the SMDE as an example.

Adjustable parameters:

AC2MODE

t.step
U.ampl
Modul.freq.
Ph.angle
Prep.cycles
Meas.cycles

SWEEP

U.start
U.end
U.step

t.drop

t

U

t

U.step

U.start

Prep.

cycles

Meas.

cycles

t.step

Ph.angle = 0

°

F (Hg)

t.modul = 1/Modul.freq.

U.ampl

× √

2

Ph.angle > 0

°

Ph.angle

Ph.angle < 0

°

90

º

180

º

270

º

360

º

Ph.angle

-180

º

-90

º

90

º

180

º

270

º

360

º

(I

n+

)

(current addition)

(I

n–

)

(current subtraction)

t.meas

(I

n+

)

t.meas

(I

n–

)

t.meas

(I

n–

)

t.meas

(I

n+

)

Area (Hg)