background image

4.2 Lead determination with standard addition

746 VA Trace Analyzer / 747 VA Stand

4-15

The dialog page "DOCUMENTATION" can be used for the entry of reports and curves
which should be put out automatically to the specified destination at the end of the
determination. A single line or all these lines can also be executed later using the
softkeyy
or .
You will find further information regarding this dialog page in section 5.6.5.

The three lines entered in the present example have the following mean-

ing:

Report Full

Printout of the full result report on the
built-in or connected printer

Curve Smth Sbst:Pb VR:**

Printout of all smoothed substance
curves on the built-in or connected printer

Curve Cal Sbst:Pb

Printout of the calibration curve (graphical
representation of the standard addition)
on the built-in or connected printer

If you do not desire an automatic printout in the subsequent lead deter-

mination, you can either delete the three entered lines with the softkey

or clear the output destination

Printer

for each line with

the key.

Press the softkey

. The subpage of "DOCUMENTATION"

appears with the method comment:

Method: XXXXXXXX DOCUMENTATION

Comment
Method: - Pb determination with DP polarography from -200 mV to -550 mV.

- This method can be used as a first test after installing the

746 VA Trace Analyzer and the 747 VA Stand or can be used as

standard operating procedure (SOP) for a GLP test of 746/747.

- For detailed information see instruction manual.

Electrodes: - Multi-Mode electrode

- Ag/AgCl reference electrode with KCl 3 mol/L

- Pt auxiliary electrode

Reagents: - KCl 3 mol/L (6.2308.020)

- Pb ion standard 1 g/L (6.2301.100)

- dist. or deion. water

Procedure: - Add 20 mL of water, 0.5 mL of KCl 3 mol/L and 100

µµL of

Pb ion standard 1 g/L (that means 100

µµg Pb)

into the well rinsed measuring vessel.

- Enter '100 uL' for 'Sample size/S0' on MONITORING page.

- Start determination.

- Add 2 * 100

µµL of Pb ion standard 1 g/L (standard addition).

Help

Delete

line

Insert

line

Get

line

Put

line

>>

Output

The dialog page "DOCUMENTATION" includes a subpage on which maximum 20
lines of free text can be entered as a comment on the method.
You will find further information regarding this dialog page in section 5.6.5.

In the present example, details of the method, the electrodes and rea-

gents to be used and the procedure used in the lead determination are

entered.

Finally press the key to return to the main page of "DOCUMEN-

TATION".