4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-38

13 Storing the calibration data in the method

The recalculated coefficients of the calibration curve are first stored only

in the method loaded in the working storage. To store these data perma-

nently, this method must be copied to the method memory. Proceed as

follows:

Press the yellow key. The dialog page "METHODS"

appears showing the methods stored in the instrument.

If the

CALCRV.mth

method is secured (an

*

is shown on the left

of the file name), it must first be unsecured. For this, select the

CALCRV.mth

method with the cursor keys. Press the softkey

< >> >

and then the softkey

. The asterisk

next to the file name disappears, the method is unsecured. Then

press the softkey

< >> >

again.

Press the softkey

. The softkey bar changes and

presents the sources for copying a method.

Press the softkey

. The

CALCRV.mth

method

is copied from the working storage to the method memory; the

method of the same name containing the old calibration data will

be overwritten.

4.3.5

Performing the determination

To perform the lead determination using the calibration curve previously recorded,

the following steps are necessary:

1

Switch to dialog page "MONITORING"

Press the yellow key. The dialog page "MONITORING"

whose header shows the loaded method

CALCRV

appears.

2

Switch the program run mode to

determination

If the parameter

Run mode

is set to

calibration

, proceed as follows:

Use the cursor keys to select the field

Run mode

.

Press the space bar to enter the edit mode.

Press the softkey

and confirm this entry with the

key.

3

Entry of the sample volume
Enter the sample volume as follows:

Use the cursor keys to select the first field below the title

Sample

size/S0

.

Press the space bar to enter the edit mode.

Enter

10 mL

and confirm this entry with the key.