Metrohm 746 VA Trace Analyzer User Manual
Page 43

3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-16
3.4.2
Filling the MME with mercury
When handling mercury, it is necessary to take special precautionary
measures. These are described in detail in section 7.2.
All actions involving the electrode and mercury vessels must be
performed in or over the drip pan 113
113 supplied (see Fig. 14).
The Hg reservoir 103
103 of the multi-mode electrode 38
38 is filled with mercury of the
highest degree of purity (mass fraction w
≥
0.99999) as follows:
1 Prepare multi-mode electrode
•
Unscrew locking ring 95
95 from slotted screw 96
96 (this gray PVC ring is
needed only to remove the slotted screws 96
96 or 101
101, see section 3.4.7
and section 3.4.9).
•
Turn slotted screw 96
96 in or out of the screw thread 98
98 using a suitable
coin until the contact surface of the black O-ring at the Plexiglas wall (thin,
black stripe) is just visible below the metal thread 98
98.
•
Remove the plastic cap used as a transport safeguard from the retaining
nut 108
108.
•
Undo retaining nut 108
108 fully and remove from screw thread 104
104.
•
Place multi-mode electrode 38
38 with the capillary opening facing upwards
in the electrode holder 114
114 (see Fig. 13).
111
111
112
112
113
113
38
38
114
114
Mercury
Fig. 13: Adding the mercury
38
38
Multi-mode electrode
(6.1246.0020)
113
113
Drip pan (6.2711.030)
111
111
Syringe (6.2816.020)
114
114
Electrode holder
(6.2615.030)
112
112
Needle (6.2816.030)