5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-234

END

Page:

O, S

Stand:

M, A

Abbreviation:

E

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## END .....

Function

End of the main program ("OP.SEQUENCE") or end of the segment
("SEGMENTS").

Description

The

END

instruction closes the main program on the "OP.SEQUENCE"

page or a segment on the "SEGMENTS" page.

Comments

Each program or segment must be closed with the

END

instruction.

The

END

instruction in the main program switches off all functions still

active at the 747 VA Stand (e.g. stirrer, purging, electrodes, voltage)
with exception of the standby voltage

U.standby

defined in a

ØMEAS

instruction.

In contrast to

END

in the main program, the

END

instruction in the

segment simply causes a return to the location in the main program
where the segment was called up.

see also

STIR, PURGE, DME, HMDE, SMDE, RDE, MEAS