beautypg.com

272 rem, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 423

background image

5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-272

REM

Page:

O, S

Stand:

M, A

Abbreviation:

RE

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. REM ..... ....................................................

Function

Remarks or comments line.

Parameter

Range; Default value / Meaning

Remark

52 characters; -

Remark

All numbers, letters and characters available on the keypad can be
entered, in addition the entry of all ASCII characters with decimal code
032

255 is possible (see section 5.2.3). The entry of the special

characters with decimal code 000

031 is not allowed as they are used

as control characters for the printer

Description

The

REM

instruction is used for the entry of remarks and comments in

programs.

Comments

If a remark should be put out on the screen as a message during the
program run, the

HOLD

instruction must be used for this.

If a remark should be printed out on the printer during the program
run, the

PRINT

instruction must be used for this.

If a remark should be printed out on an external printer attached to
one of the two RS232 interfaces, the

SEND>RS1

or

SEND>RS2

instruc-

tion must be used.

see also

HOLD, PRINT, SEND>RS1, SEND>RS2