5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-193

Example

Addition of a standard addition solution with the 665 Dosimat via the
remote interface:

746 connection:

Connection "Control Lines"

23

23

(see section 3.13)

665 connection:

Connection

88

for external dosing contact

(see 665 Instructions for Use)

Connecting cable: Cable with 25-pin connector, pin 5 (output 1) connected to red

banana plug, pin 25 (0 V) to black banana plug

Setting on 665:

"DIS R" mode with fixed dispensing volume

V.add

Program:

:

SMPL>M or SMPL/M

Sample addition to measuring vessel

(ADD

Loop start for standard additions

(REP

Loop start for replication

SEGMENT

Call-up of the segment (measurement)

REP)n

Loop end, n = number of replications

SETCTRL *******1

Start of dispensing at 665 Dosimat

:

by setting control line 1 active

NOP xxx s

Entry of a wait time until

dispensing

:

is at an end

SETCTRL *******0

Resetting of control line 1

ADD/M

Standard addition solution has been added

ADD)n

Loop end, n = number of standard additions

:

see also

(ADD

,

ADD)n

,

ADD>M

,

PADD>M

,

SETCTRL

,

SEND>RS#