Metrohm 746 VA Trace Analyzer User Manual
Page 398

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-247
Examples
Determination of cadmium by stripping voltammetry:
Method ASV-1 .mth OPERATION SEQUENCE
Title Determination of cadmium
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 SMPL>M V.fraction mL V.total L
2 PURGE
3 STIR 300.0 Rot speed 2000 /min
4 (REP
5 PURGE
6 STIR 20.0 Rot speed 2000 /min
7 ØPURGE 5.0
8 SEGMENT Segm.name CdASV
9 REP)2
.
.
10 END
.
.
Method ASV-1.mth SEGMENT
CdASV
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 HMDE Drop size 4 Meas.cell normal
2 DPMODE U.ampl 50 mV t.meas 20.0 ms
t.step 0.20 s t.pulse 40.0 ms
3 MEAS 60.0 U.meas -850 mV
4 ØSTIR 5.0
5 SWEEP 14.0 U.start -850 mV U.step 6 mV
U.end -450 mV Sweep rate 30 mV/s
6 END
.
In the
CdASV
segment, cadmium is enriched at the HMDE for 60 s at
–850 mV and then determined in a sweep from –850 mV ... –450 mV.
see also
ØMEAS, HMDE, TSWEEP, END