beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 398

background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-247

Examples

Determination of cadmium by stripping voltammetry:

Method ASV-1 .mth OPERATION SEQUENCE

Title Determination of cadmium

================================================================================

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

1 SMPL>M V.fraction mL V.total L

2 PURGE

3 STIR 300.0 Rot speed 2000 /min

4 (REP

5 PURGE

6 STIR 20.0 Rot speed 2000 /min

7 ØPURGE 5.0

8 SEGMENT Segm.name CdASV

9 REP)2

.

.

10 END

.

.

Method ASV-1.mth SEGMENT

CdASV

================================================================================

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

1 HMDE Drop size 4 Meas.cell normal

2 DPMODE U.ampl 50 mV t.meas 20.0 ms

t.step 0.20 s t.pulse 40.0 ms

3 MEAS 60.0 U.meas -850 mV

4 ØSTIR 5.0

5 SWEEP 14.0 U.start -850 mV U.step 6 mV

U.end -450 mV Sweep rate 30 mV/s

6 END

.

In the

CdASV

segment, cadmium is enriched at the HMDE for 60 s at

–850 mV and then determined in a sweep from –850 mV ... –450 mV.

see also

ØMEAS, HMDE, TSWEEP, END