Metrohm 746 VA Trace Analyzer User Manual
Page 289

5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-138
Parameter
Range; Default value / Meaning
AC2 abs.
Linear baseline with absolute evaluation with
AC2MODE
:
Curves recorded with AC2MODE are always evaluated
with a linear baseline. The base points
dU.front
and
dU.rear
can be determined automatically or entered
manually, the slopes
S.front
and
S.rear
have no
meaning here.
In the absolute evaluation, the total peak height
I.abs
is
determined and put out as the result.
In place of
AC2 abs.
,
linear
can be entered under
Type
.
I.abs
U
/
2
AC2 abs.
dU.front dU.rear
Scope
whole, f.half, r.half, f.double, r.double; whole
Baseline range
The following possibilities exist to define the range over which the
baseline should be calculated:
whole
Calculation of the baseline for the whole curve.
whole
f.half
Calculation of the baseline starting from peak front half
(possible only with
Type = linear
).
r.half
Calculation of the baseline starting from peak rear half
(possible only with
Type = linear
).
With
f.half
and
r.half
is recommended to set
S.front
and
S.rear
from
auto
to
0.000
(zero).
linear, f.half
linear, r.half
f.double
Calculation of the baseline for the first peak of a double
peak (possible only with
Type = linear
,
Type =
polynom
and
Type = expon.
). The second peak must be
evaluated with the same
Type
.
r.double
Calculation of the baseline for the second peak of a
double peak (possible only with
Type = linear
,
Type =
polynom
and
Type = expon.
). The first peak must be
evaluated with the same
Type
.