5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-138

Parameter

Range; Default value / Meaning

AC2 abs.

Linear baseline with absolute evaluation with

AC2MODE

:

Curves recorded with AC2MODE are always evaluated
with a linear baseline. The base points

dU.front

and

dU.rear

can be determined automatically or entered

manually, the slopes

S.front

and

S.rear

have no

meaning here.
In the absolute evaluation, the total peak height

I.abs

is

determined and put out as the result.
In place of

AC2 abs.

,

linear

can be entered under

Type

.

I.abs

U

/

2

AC2 abs.

dU.front dU.rear

Scope

whole, f.half, r.half, f.double, r.double; whole

Baseline range

The following possibilities exist to define the range over which the
baseline should be calculated:

whole

Calculation of the baseline for the whole curve.

whole

f.half

Calculation of the baseline starting from peak front half
(possible only with

Type = linear

).

r.half

Calculation of the baseline starting from peak rear half
(possible only with

Type = linear

).

With

f.half

and

r.half

is recommended to set

S.front

and

S.rear

from

auto

to

0.000

(zero).

linear, f.half

linear, r.half

f.double

Calculation of the baseline for the first peak of a double
peak (possible only with

Type = linear

,

Type =

polynom

and

Type = expon.

). The second peak must be

evaluated with the same

Type

.

r.double

Calculation of the baseline for the second peak of a
double peak (possible only with

Type = linear

,

Type =

polynom

and

Type = expon.

). The first peak must be

evaluated with the same

Type

.