3.6 Reference electrode

746 VA Trace Analyzer / 747 VA Stand

3-29

3.6

Reference electrode

3.6.1

Construction

The complete reference electrode (RE) 43

43 comprises two parts:

6.0728.0X0

Ag/AgCl reference system (122

122)

with ceramic diaphragm type D, diameter = 1 mm
6.0728.020 Reference system: Ag/AgCl/c(KCl) = 3 mol/L;

supplied in a holder filled with c(KCl) = 3 mol/L

as standard

6.0728.010 Reference system: Ag/AgCl

supplied dry (option)

6.1245.010

Electrolyte vessel (123

123)

with ceramic diaphragm type D, diameter=3 mm;

holds a second electrolyte solution (bridge

electrolyte) and thus forms with the 6.0728.020

Reference system a complete reference electrode

in the so-called double junction construction.

The construction of the reference electrode and the designations of the individual

parts are shown in Fig. 17.

43

123

130

131

132

122

124

125
126
127
128

129

Fig. 17: Construction of the reference electrode

43

43

Reference electrode

122

122

Reference system (6.0728.0X0)

123

123

Electrolyte vessel (6.1245.010)

124

124

Electrical connection for cable

"AE"

125

125

Vent opening

126

126

Ag/AgCl reference system

127

127

Electrolyte compartment with

internal electrolyte

128

128

Diaphragm support made of

PCTFE

129

129

Diaphragm

130

130

Vent opening

131

131

Electrolyte compartment with

bridge electrolyte

132

132

Diaphragm