5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-164

Parameter

Range; Default value / Meaning

Operators
The operators are processed from left to right. With nested expres-
sions, multiplication and division have precedence over addition and
subtraction.

=

Equals (value assignment)

+

Addition

Subtraction

*

Multiplication

/

Division

( )

Brackets
Max. 4 pairs of brackets can be nested.

Standard formulae
Pressing the softkey

enters standard formulae in

this field which provide the mass concentration of the substances
referred to the sample size

S0

as a final result in g/L or g/g.

Automatic entry of the standard formulae depends on whether the
sample size has been entered as a weight or volume and whether
diluted or undiluted samples have been used (see also section 5.4.4).
The dependence of the put in sample size

S0

and the parameters of

the SMPL instructions is as follows:

S0 S2

V.fraction

V.total

Standard formula for final

result

Result

unit

L

MC:

g/L

g

d

MC:

/

(S2

*

1000)

g/g

L

V.fraction

V.total

MC:

*

V.tot

/

S0

g/L

g

V.fraction

V.total

MC:

*

V.tot

/

S0

g/g

The standard formulae listed above refer to the basic quantities g and
L, but the results are automatically supplied with the appropriate prefix
by the simultaneous entry of

#g/L

or

#g/g

in the field

Res.unit

(see

Res.unit

).

Standard formulae can be altered at a later date to specify the result in
another unit (e.g. ppm) or for further calculation.

In the case of modifications to the method which also

influence the standard formulae, any existing standard

formulae must be deleted and then regenerated with the

softkey

).