5.6.3 SUBSTANCES

746 VA Trace Analyzer / 747 VA Stand

5-149

Parameter

Range; Default value / Meaning

Determining working range

Extrapolations above the range determined with calibration solu-
tions are not allowed. Thus, enter the maximum admissible
working range on the dialog page "SUBSTANCES".

Keep temperature constant

Owing to the large temperature dependence of the measured
values (

2%/°C), it is advisable to work with the 6.1418.220 ther-

mostatted measuring vessel.

Consideration of blank values

Any blank values must be determined separately and subtracted
in the formula calculation on the dialog page "CALCULATIONS".

Curve type

linear, 0/linear, nonlin, 0/nonlin; linear

Curve type for calibration curve

The calibration curve can be determined with linear or nonlinear
functions:

linear

Straight line of type y = a + bx
Calculation by weighted linear regression.
Minimum number of calibration solutions: 2

0/linear

Straight line of type y = bx
Calculation by weighted linear regression, the line is
passed through the origin.
Minimum number of calibration solutions: 1

nonlin

Curve of type y = a + bx + cx

4

Calculation by weighted curve matching.
Minimum number of calibration solutions: 3

0/nonlin

Curve of type y = bx + cx

4

Calculation by weighted curve matching, the curve is
passed through the origin.
Minimum number of calibration solutions: 2

y:

EV

Evaluation value (A, W or s)

x:

ρ

eff

Effective mass concentration in the
analysis solution (g/L)

a:

Y.offset

Intercept

b:

Slope

Slope of the line or slope of the curve
in the linear region

c:

Nonlin.

Non-linearity factor