5.3.2 STAND

746 VA Trace Analyzer / 747 VA Stand

5-25

Parameter

Range; Default value / Meaning

Pressure

on, off; off

Inert gas pressure (see

)

Purging

on, off; off

Purging (see

)

Stirrer

on, off; off

Stirrer (see

)

DME

on.dt, off; off

DME (Dropping Mercury Electrode) (see

)

HMDE

on, off; off

HMDE (Hanging Mercury Drop Electrode) (see

)

SMDE

on.dt, off; off

SMDE (Static Mercury Drop Electrode) (see

)

RDE

on, off; off

RDE (Rotating Disk Electrode) (see

)

Rot.speed

0, 200, 400, ..., 3000 /min; 2000 /min

Rotational speed of stirrer

Setting of the rotational speed of the stirrer for manual operation
(revolutions/min).

Drop size

1...9; 5

Drop size of the HMDE

Setting of the drop size of the HMDE (Hanging Mercury Drop Elec-
trode) for manual operation.

The surface area of the mercury drop is 0.15...0.60 mm

2

, depending

on the setting.

Drop size

1...9; 5

Drop size of the SMDE

Setting of the drop size of the SMDE (Static Mercury Drop Electrode)
for manual operation.

The surface area of the mercury drop is 0.15...0.60 mm

2

, depending

on the setting.

The time for formation of the drop is:

Drop time

t.drop

=

Drop size

×

40 ms

Rot.speed

0, 200, 400, ..., 3000 /min; 2000 /min

Rotational speed of RDE

Setting of the rotational speed of the RDE (Rotating Disk Electrode) for
manual operation (revolutions/min).