4.3 Lead determination with calibration curve

746 VA Trace Analyzer / 747 VA Stand

4-37

The full report

Full report

essentially contains the same parts as that

of the

TESTPb.mth

method with standard addition (see section 4.2.6).

However, the following sections are different:

1

Data for substance

Only the substance name

Pb is given, the other

evaluation

parameters remain blank as they are not calcula-
ted when the calibration curve is recorded.

2

Data for calibration

Offset

Y.offset and slope Slope of the

curve

calibration curve.

3

Final results

The final result is not calculated as the variable
MC:Pb is not calculated when the calibration
curve is recorded.

Finally, the calibration curve is printed out:

================== METROHM 746 VA TRACE ANALYZER (5.746.0101) ==================

Method : CALCRV User : Date : 1996-12-15

Mode : DPMODE Determ. : 12151548 Time : 15:48:08

Run : 2 Sample :

Calibration curve

Curve type : lin Slope : 1.438e-03 Mean.dev. : 5.933e-10

Y.reg : 5.489e-10

Substance : Pb

The printout of the calibration curve essentially contains the same parts

as the

TESTPb.mth

method with standard addition (see section 4.2.6).

However, the following sections are different:

1

Data for calibration curve Curve type, addition mode, slope of the calibration

curve, offset

Y.offset of the calibration curve,

mean deviation of the measured values about the
calibration curve.

2

Calibration curve

Measured points (mean values of the measured
peak currents for the calibration solutions),
calculated calibration curve, current axis, axis for
effective mass concentration in the measuring
vessel.

2

1