Metrohm 746 VA Trace Analyzer User Manual
Page 122

4.3 Lead determination with calibration curve
746 VA Trace Analyzer / 747 VA Stand
4-37
•
The full report
Full report
essentially contains the same parts as that
of the
TESTPb.mth
method with standard addition (see section 4.2.6).
However, the following sections are different:
1
Data for substance
Only the substance name
Pb is given, the other
evaluation
parameters remain blank as they are not calcula-
ted when the calibration curve is recorded.
2
Data for calibration
Offset
Y.offset and slope Slope of the
curve
calibration curve.
3
Final results
The final result is not calculated as the variable
MC:Pb is not calculated when the calibration
curve is recorded.
•
Finally, the calibration curve is printed out:
================== METROHM 746 VA TRACE ANALYZER (5.746.0101) ==================
Method : CALCRV User : Date : 1996-12-15
Mode : DPMODE Determ. : 12151548 Time : 15:48:08
Run : 2 Sample :
Calibration curve
Curve type : lin Slope : 1.438e-03 Mean.dev. : 5.933e-10
Y.reg : 5.489e-10
Substance : Pb
•
The printout of the calibration curve essentially contains the same parts
as the
TESTPb.mth
method with standard addition (see section 4.2.6).
However, the following sections are different:
1
Data for calibration curve Curve type, addition mode, slope of the calibration
curve, offset
Y.offset of the calibration curve,
mean deviation of the measured values about the
calibration curve.
2
Calibration curve
Measured points (mean values of the measured
peak currents for the calibration solutions),
calculated calibration curve, current axis, axis for
effective mass concentration in the measuring
vessel.
2
1