5.3 Set Up
746 VA Trace Analyzer / 747 VA Stand
5-22
Softkeys
Meaning
Stirrer
Switch stirrer on/off
The stirrer is switched on or off. The current status is shown in the
Stirrer
parameter.
Even when the "STAND" page is quit, the stirrer remains switched on
until it is switched off manually or automatically (by a program start).
DME
Switch DME (Dropping Mercury Electrode) on/off
The DME is switched on or off. After switching on, the mercury drops
freely from the MME for ca. 1 min. This is shown in the
DME
parameter
by
on.dt
.
After the "STAND" page has been quit or on change to another type of
electrode, a DME which is switched on is automatically switched off.
HMDE
Switch HMDE (Hanging Mercury Drop Electrode) on/off
The HMDE is switched on or off. On switching on, a new mercury drop
is formed at the MME. This is briefly shown in the
HMDE
parameter by
on
.
The size of the mercury drop can be set with the
Drop size
parame-
ter.
SMDE
Switch SMDE (Static Mercury Drop Electrode) on/off
The SMDE is switched on or off. After switching on, mercury drops
from the MME in a controlled manner for ca. 1 min. This is shown in
the
SMDE
parameter by
on.dt
.
The size of the mercury drop can be set with the
Drop size
parame-
ter.
RDE
Switch RDE (Rotating Disk Electrode) on/off
The RDE is switched on or off. The current status is shown in the
RDE
parameter.
>>
Settings
Switch to Settings
Switches the softkey bar to the settings parameters and other func-
tions.
VA Stand
1
Switch to VA Stand 1
Maximum three 747 VA stands can be attached to the 746 VA Trace
Analyzer. This softkey is used to switch to the VA Stand with stand
address
1
.
The stand currently selected is displayed in the title line:
Manual Address 1
VA Stand with address 1
Manual Address 2
VA Stand with address 2
Automatic Address 3
VA Stand for Autosampler with address 3
VA Stand
2
Switch to VA Stand 2
Maximum three 747 VA stands can be attached to the 746 VA Trace
Analyzer. This softkey is used to switch to the VA Stand with stand
address
2
.
The stand currently selected is displayed in the title line:
Manual Address 1
VA Stand with address 1
Manual Address 2
VA Stand with address 2
Automatic Address 3
VA Stand for Autosampler with address 3