282 send>rs2, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 433: Send>rs2

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-282
SEND>RS2
Page:
O, S
Stand:
M, A
Abbreviation:
SEND>RS2
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SEND>RS2 ..... ....................................................
Function
Send a message to RS interface 2.
Parameter
Range; Default value / Meaning
Remark
52 characters; -
Message to be sent
The remark entered here is sent to RS232 interface 2.
All numbers, letters and characters on the keypad can be entered. The
following entry possibilities also exist:
Щ
Щ
###
Characters with decimal code
###
(see section
5.2.3). Characters with decimal code 000 ... 031
should not be entered as, with the exception of 010
(LF), although they can be shown on the screen, they
can not be printed out.
ЩЩ
ЩЩ
Character
"ЩЩ"
Щ
Щ
092###
Characters with octal code
###
(
Щ
Щ
092
=
"\"
)
Щ
Щ
092x##
Characters with hex code
##
Щ
Щ
092r
Carriage return (CR)
Щ
Щ
092n
Line feed (LF)
Щ
Щ
092f
Form feed (FF)
Щ
Щ
092b
Backspace (deletion of the character on the left)
Щ
Щ
092t
Tab (tabulator)
Щ
Щ
092ЩЩ092
Character
"\"
At the end of the transmitted line, there is no
automatic CR/LF. This has to be explicitly entered
(e.g. with "
........ ÙÙ092r ÙÙ092n
").
Description
The
SEND>RS2
instruction is used to send the message entered in the
parameter Remark to RS232 interface 2 or output it on a device attached
to this interface. It is thus possible to send, e.g. control commands to an
external device.
Comments
For the printout of a message on the built-in thermal printer or the
external printer connected to the parallel interface, the
instruction
must be used.
see also
SEND>RS1, PRINT