2 installation of 746 va trace analyzer, Setting the instrument supply voltage -2, Fuses -2 – Metrohm 746 VA Trace Analyzer User Manual
Page 29: Mains cable and mains connection -3, Switching the instrument on/off -3, Installing paper (only for internal printer) -3, Handling of data cards -4

3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-2
3.2
Installation of 746 VA Trace Analyzer
Follow the instructions below for connecting to the power supply. If
the instrument is operated with the mains voltage set wrongly and/or
wrong mains fuse there is a danger of fire!
3.2.1
Setting the instrument supply voltage
Before switching on the 746 VA Trace Analyzer for the first time, check that the
mains voltage set on the instrument (visible in mains voltage selector 99) matches
the local power supply voltage. If this is not the case, you must change the voltage
set on the instrument by moving the mains voltage selector 99 with a screwdriver:
230V
115V
3.2.2
Fuses
The 746 VA Trace Analyzer contains 2 fuses 10
10 and 12
12 built in as standard, both of
type 2 ATH (2 A, slow-blow, high breaking capacity, Metrohm ordering No.
U.600.0107).
Ensure that the instrument is never put into operation with fuses of
another type, otherwise there is danger of fire!
To change blown fuses, proceed as follows
1
Change Fuse 10
10 (neutral, 2 ATH, slow-blow)
•
Disconnect mains cable from mains connection plug 13
13.
•
Using a screwdriver, turn the fuse holder 10
10 to the left until it can be
pulled out.
•
Take out fuse holder, remove blown fuse and insert replacement fuse
(2 ATH, slow-blow) in the holder.
•
Reinsert fuse holder in instrument, push in with a screwdriver and fix by
turning to the right.
2
Change Fuse 12
12 (phase, 2 ATH, slow-blow)
•
Disconnect mains cable from mains connection plug 13
13.
•
Using a screwdriver, push the fuse holder 12
12 upwards until it can be
pulled out.
•
Take out fuse holder, remove blown fuse and insert replacement fuse
(2 ATH, slow-blow) in the holder.
•
Push fuse holder back into instrument until it clicks into place.
Positions of mains voltage
selector 99:
115: 100 ... 120 V ± 10%
230: 220 ... 240 V ± 10%