3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-2

3.2

Installation of 746 VA Trace Analyzer

Follow the instructions below for connecting to the power supply. If

the instrument is operated with the mains voltage set wrongly and/or

wrong mains fuse there is a danger of fire!

3.2.1

Setting the instrument supply voltage

Before switching on the 746 VA Trace Analyzer for the first time, check that the

mains voltage set on the instrument (visible in mains voltage selector 99) matches

the local power supply voltage. If this is not the case, you must change the voltage

set on the instrument by moving the mains voltage selector 99 with a screwdriver:

230V

115V

3.2.2

Fuses

The 746 VA Trace Analyzer contains 2 fuses 10

10 and 12

12 built in as standard, both of

type 2 ATH (2 A, slow-blow, high breaking capacity, Metrohm ordering No.

U.600.0107).

Ensure that the instrument is never put into operation with fuses of

another type, otherwise there is danger of fire!

To change blown fuses, proceed as follows

1

Change Fuse 10

10 (neutral, 2 ATH, slow-blow)

Disconnect mains cable from mains connection plug 13

13.

Using a screwdriver, turn the fuse holder 10

10 to the left until it can be

pulled out.

Take out fuse holder, remove blown fuse and insert replacement fuse

(2 ATH, slow-blow) in the holder.

Reinsert fuse holder in instrument, push in with a screwdriver and fix by

turning to the right.

2

Change Fuse 12

12 (phase, 2 ATH, slow-blow)

Disconnect mains cable from mains connection plug 13

13.

Using a screwdriver, push the fuse holder 12

12 upwards until it can be

pulled out.

Take out fuse holder, remove blown fuse and insert replacement fuse

(2 ATH, slow-blow) in the holder.

Push fuse holder back into instrument until it clicks into place.

Positions of mains voltage
selector 99:

115: 100 ... 120 V ± 10%

230: 220 ... 240 V ± 10%