3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-22

10 Pressurize the MME

Switch on 746 VA Trace Analyzer with mains switch 11

11 (746 VA Trace

Analyzer and 747 VA Stand must first be installed properly as described

in section 3.2 and section 3.3).

Select dialog page "STAND" and then switch on the inert gas supply to

the 747 VA Stand by pressing the softkey

. This pressurizes

the multi-mode electrode 38

38 and the mercury begins to flow slowly out of

the capillary.

Gently tap the MME with your finger (to remove any air bubbles) and

allow mercury to flow out of the capillary into the empty measuring vessel

for approx. 2 min.

Fill measuring vessel 28

28 with 10 mL ultrapure water and add 1 drop KCl

solution (in pure water, mercury drops from the capillary only with diffi-

culty).

Allow mercury to flow out of the capillary for ca. 2 min while checking the

drop formation: The drop time should be ca. 3 s.

11 Adjusting the sealing needle 97

97

Turn slotted screw 96

96 using a suitable coin slowly in a clockwise direction

until the mercury flow stops.

Open slotted screw 96

96 slightly in an anticlockwise direction until the

mercury flow restarts.

Gently tap the MME with your finger and turn the slotted screw 96

96 very

slowly clockwise until the mercury flow just stops. (The tapping action is

used to knock off the mercury drop so that it is easier to see whether

mercury continues to flow.)

Finally, turn slotted screw 96

96 a quarter of a turn clockwise.

12 Checking the MME for leaks

Select dialog page "STAND" and then switch on the dropping mercury

electrode by pressing the softkey

. The mercury drops freely out of

the capillary for ca. 1 min.

Press softkey

. A single mercury drop is formed. Knock this off by

gently tapping the MME 38

38 with your finger and check that the mercury

flow has really stopped. Repeat this operation several times.

If mercury continues to flow, turn slotted screw 96

96 still further in a clock-

wise direction and repeat check.

3.4.6

Storing the MME

On completion of the measurements, the MME is stored in the 747 VA Stand so

that the tip of the glass capillary 109

109 is immersed in pure water (or in the solvent

used). This prevents blockage of the capillary by crystallized salts.
An electrode treated in this manner can be taken out of the 747 VA Stand after a

few hours and stored in air for a lengthy period without suffering any damage. Al-

ways store the MME so that the glass capillary 109

109 is horizontal.