beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 145

background image

4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-60

6

Addition of acetate buffer

Use a pipette to add 1 mL acetate buffer (1 mol/L, pH 4.6) to the

measuring vessel through the pipetting opening. Then close the pipet-

ting opening using stopper 25

25 and press the key.

7

Measurement of the sample solution

Measurement of the sample solution runs as follows (see also section

4.5.3):

Purging and stirring for 330 s.

Wait time 5 s to stabilize the analysis solution.

Call-up of the segment

polaro

for measured value recording.

The single measurement is allocated the VR number

00

.

On completion of the measurement, the program is at an end. The status

display in the header of the dialog page "MONITORING" switches back

to

* READY *

, the instrument is ready for a new determination.

8

Printout of the segment curve

On completion of the program, the smoothed segment curve entered on

the "DOCUMENTATION" page is put out on the printer. The following ex-

ample shows a determination with drinking water as sample:

================== METROHM 746 VA TRACE ANALYZER (5.746.0101) ==================

Method : POL User : Date: 1996-12-10

Mode : DPMODE Determ. : 12101420 Time: 14:20:12

Run : 6

Curve : smoothed Segment : polaro V R : 0 0