7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-9

Error message

Meaning/Causes

Rectification

Exit

CALIBR - ERROR 6:
Missing reference

No reference to reference

substance in the calibration

with internal standard (no

evaluation possible)

No reference substance

entered

No result for reference

substance found

Enter reference substance on the

”SUBSTANCES/Calibration” page

Check peak recognition for reference

substance

E/C

CALIBR - ERROR 7:
Sample mass too low

Sample mass in measuring

vessel <0.1 pg (no evaluation

possible)

Check sample size

S0

on the

”SAMPLES” page

Check program, above all the

SMPL/M

and

SMPL>M

instructions

E/C

CALIBR - ERROR 8:
Measuring volume out of
range

Solution volume <1 mL or

>

V.meas max

Check sample size

S0

on the

”SAMPLES” page

Check program, above all the

SMPL/M

and

SMPL>M

instructions

Change

V.meas max

E/C

CALIBR - ERROR 9:
Missing evaluation
values

Too few measured points

(calibration solutions) for the

selected calibration mode

No calibration curve available

No results for calibration

curve found

Quantity = P.peak

inadmissible for

AC2MODE

Quantity = P.peak

inadmissible for

DCTMODE

with

SWEEP

Increase number of measured points

(calibration solutions) or change

calibration mode

Plot new calibration curve or – if no

calibration desired – set

Technique

to

none

Check peak recognition for calibration

Set

Quantity

to

I.peak

Set

Quantity

to

I.peak

or use

DCTMODE

with

SSWEEP

E/C

CONFIG - ERROR 1:
Bad Autosampler
configuration

Stand address and configu-

ration entry on the ”VA

TRACE ANALYZER” page do

not agree

Wrong configuration of 695

Autosampler

Change stand address to

3

or

restart VA Trace Analyzer

Install Autosampler according to

section 3.9

E/S

CONFIG - ERROR 2:
Illegal Dosimat number

Wrong Dosimat number

Set Dosimat number to 1...4

E/S

CONFIG - ERROR 3:
Illegal stand number

Wrong stand number

Set stand number to

1

or

2

E/S

CONFIG - ERROR 4:
Bad VA Stand
configuration

Stand address and configura-

tion entry on the ”VA TRACE

ANALYZER” page do not agree

Change stand address or restart VA

Trace Analyzer

E/C

CONFIG - ERROR 5:
Bad Dosimat
configuration

Dosimat address and configu-

ration entry on the ”VA TRACE

ANALYZER” page do not agree

Change Dosimat address or restart VA

Trace Analyzer

E/S

CONFIG - ERROR 6:
Bad Sample Changer
configuration

Address of the 717 Sample

Changer and configuration

entry on the ”VA TRACE

ANALYZER” page do not agree

Change address at 717 Sample

Changer or restart VA Trace Analyzer

E/S

CURVES - ERROR 1:
No data assigned to
substance

No result data for selected

substance available

Perform recalculation

If error reappears, perform initializa-

tion (see section 7.3.3)

If error reappears, start new meas-

urement

E

CURVES - ERROR 2:
Modify baseline with
smoothed curve only

Modify baseline

is possible

only with smoothed or sub-

tracted curves

Switch to smoothed or subtracted curve